Strive for 5 vaccine fridge temperature chart

Record the current, minimum and maximum temperature of a vaccine or medicine fridge twice a day, reset the thermometer, and document any reading outside 2°C to 8°C.

Free to use and adapt. Includes blank records, instructions and fictional examples.

Version 1.0 · Source guidance reviewed 18 September 2026
Pharmacy · General practice · Immunisation clinics · Aged care · Veterinary

Actual blank twice-daily vaccine fridge temperature chart form included in this pack

A4 landscape PDF. Select the preview to open the full document.

How to use this template

  1. Write the practice or pharmacy name, the fridge identifier and the month at the top. Use one chart per fridge and per month.
  2. Twice a day, at the start and end of the working day, read the current, minimum and maximum temperatures from the fridge thermometer or data logger and record them to one decimal place.
  3. Reset the minimum and maximum after each reading and tick the reset column. Initial every entry so the person who checked can be identified.
  4. If any reading is outside 2°C to 8°C, isolate the vaccines, mark them do not use, do not discard, and contact your state or territory immunisation program for advice before using or disposing of stock. Record what you did in the action column.
  5. Plot the morning and afternoon readings on the 2°C to 8°C band on page 3 to see drift across the month, and keep completed charts for the period your accreditation program requires.

Inside the pack

Use the blank forms for your own records. The examples below show how fields are used; they are fictional and do not prescribe operational limits.

Twice-daily vaccine fridge temperature chart

One row per day. Record the morning and afternoon readings, reset the minimum and maximum each time, and note any action taken for a reading outside 2°C to 8°C.

See fields and worked example
Date
14/10/2026
AM current °C
4.8
AM minimum °C
3.9
AM maximum °C
5.6
AM reset and initials
Reset · JT
PM current °C
5.1
PM minimum °C
4.6
PM maximum °C
5.9
PM reset and initials
Reset · AS
Action if outside 2 to 8°C
Within range; no action

Before you use it

The National Vaccine Storage Guidelines (Strive for 5) ask providers to store vaccines between 2°C and 8°C, aim for 5°C, and record the current, minimum and maximum temperatures at least twice each working day, resetting the thermometer each time. A continuous data logger is also expected. This chart supports that routine; it does not replace the guideline, your cold chain breach protocol or the official poster.

Adapt these supporting records to your products, processes and approved food safety plan. They are not HACCP certification or a complete site-specific food safety program. Establish and validate limits with a competent person before use. Examples are fictional and are not prescribed limits.